Medium-Temperature Area Source Blackbody
Z4MHT-Ø165 / Z5MHT-170
Product Image: Medium-Temperature Area Source Blackbody
Product Overview
As a calibration blackbody, the medium-temperature area source blackbody is calibrated with standard instruments from the Metrology Institute and can stably provide blackbody radiation with equivalent temperature for real-time temperature calibration during the temperature measurement process of infrared thermal imaging temperature measurement systems. It features a wide operating range, large radiation area, short constant temperature time, and good temperature stability.
Product Features
- The blackbody radiation surface is sandblasted and coated with high emissivity coating, which does not fade or peel at high temperatures;
- The radiation surface features a pyramid design with high effective emissivity;
- High temperature control accuracy and good temperature stability;
- The heating part is separated from the outer shell and equipped with a heat dissipation air duct, effectively reducing the outer shell temperature;
- Calibrated with standard instruments from the Metrology Institute to ensure temperature accuracy;
- Multiple temperature points can be corrected to ensure temperature accuracy.
Technical Parameters
| Model | Z4MHT-Ø165 | Z5MHT-170 |
|---|---|---|
| Model: Temperature Setting Range | Z4MHT-Ø165: Room Temperature +5℃ to 400℃ | Z5MHT-170: 50 to 500℃ |
| Model: Effective Radiation Area | Z4MHT-Ø165: Ø165mm | Z5MHT-170: 170mm×170mm |
| Model: Effective Emissivity | Z4MHT-Ø165: 0.98±0.2 | Z5MHT-170: 0.98±0.02 |
| Model: Temperature Resolution | Z4MHT-Ø165: 0.1℃ | Z5MHT-170: 0.1℃ |
| Model: Temperature Stability | Z4MHT-Ø165: ≤0.1℃/10min | Z5MHT-170: ≤0.1 to 0.2℃ |
| Model: Temperature Uniformity | Z4MHT-Ø165: ≤0.5℃ (@100℃); ≤1℃ (@300℃) | Z5MHT-170: ≤0.5℃(@100℃), ≤2℃(@300℃) |
| Model: Dimensions | Z4MHT-Ø165: 185×420×365mm (Blackbody); 330×200×152mm (Controller) | Z5MHT-170: 170×380×333mm (Blackbody); 226×136×145mm (Controller) |
| Model: Weight | Z4MHT-Ø165: 14.2Kg (Blackbody); 6.8Kg (Controller) | Z5MHT-170: 15Kg (Blackbody); 2Kg (Controller) |
| Model: Operating Environment Temperature/Humidity | Z4MHT-Ø165: 25±5℃/≤80%RH | Z5MHT-170: 25±5℃/≤80%RH |
| Model: Power Supply | Z4MHT-Ø165: 220VAC/50Hz/1000W | Z5MHT-170: 220VAC/50Hz/1000W |
| Model: Communication Interface | Z4MHT-Ø165: RS232 | Z5MHT-170: RS232 |